Description
| Ventilation modes: | P-CMV, V-CMV, P-SIMV, V-SIMV, PSV, SIGH, MANUAL |
| Ventilator parameter ranges | |
| Flow meter | O2(0.1-10L/min) |
| N2O(0.1-10L/min) | |
| AIR(0.1-10L/min) | |
| Rapid oxygen supply | 35L/min-75L/min |
| Tidal volume(Vt) | 0, 20 mL ~ 1500 mL |
| Frequence(Freq) | 1bpm ~ 100 bpm |
| I/E | 4:1~1:8 |
| PEEP | 0cmH2O ~ 30 cmH2O |
| Pressure triggering sensitivity(PTr) | -20 cmH2O ~ 0 cmH2O(Based on PEEP) |
| Flow trigger sensitivity (FTr) | 0.5 L/min ~ 30 L/min |
|
Pressure control (PC) |
5 cmH2O ~ 60 cmH2O |
| SIGH | 0(off) 1/100 ~ 5/100 |
| Apnea Ventilation | OFF,5 s ~ 60 s |
| Pressure Limit | 20 cmH2O ~ 100 cmH2O |
| Monitored Parameters | |
| Frequence(Freq) | 0 /min ~ 100 /min |
| Tidal volume(Vt) | 0 mL ~ 2000 mL |
| MV | 0 L/min ~ 100 L/min |
| Oxygen concentration | 15 % ~ 100 %
|
| Graphical display: | |
| P-T(pressure – time) | |
| F-T(flow – time) | |
| P-V loop (pressure – volume loop) | |
| V-T(volume – time ) | |
|
Alarm and protection |
|
| The AC power failure alarm | Power failure or no connection |
| Internal battery backup low voltage alarm | <11.3±0.3V |
| No tidal volume | ≤5Ml within 6s |
| High oxygen concentration alarm
Low oxygen concentration alarm |
19%-100%
18%-99% |
| Suffocation warning | 5s-60s no spontaneous ventilation |
| The maximum limited pressure | <12.5 kPa |
| Fan error
Oxygen deficit
|
Show on screen
Show on screen |
| Working conditions | |
| Gas source | O2,N2O,AIR |
| Pressure | 280kPa-600kPa |
| Voltage | 100-240V |
| Power frequency | 50/60Hz |
| Input power | 80VA |
| Vaporizer | Halothane, Enflurane, Isoflurane, Sevoflurane |